NucleOSS workflow sheet · Bridge direct to implants, digital workflow

Screw-retained bridge on NO-HEX Ti-Bases

NO-HEX straight or angled screw-channel Ti-Bases for a connected bridge, with matched screws, drivers and IFU torque.

A parts reference for dental laboratories and dentists, compiled from the manufacturer documents listed below. It is not a clinical recommendation: the clinician selects the components, and the current instructions for use apply. Torque is shown only where an IFU states it. Confirm with HealthIN marks anything the cited sources do not state or state differently.

To print or save this sheet as a PDF, use your browser’s Print command.

Ask HealthIN about this sheet
  1. 1Scan

    Implant-level scan body

    One per implant. T6 32895 is the short version.

    Cited Catalogue 2026 PDF p. 27 (printed 26); D06.001/EN PDF p. 2, 3 (printed 2-3)

    Screw

    The manual names T6 16000 with T6 32899; confirm the screw for the short T6 32895 with HealthIN.

    • SM.05 PDF p. 188 (printed 179) · Digital solutions, intraoral scanT6 16000 Abutment Screw · T6 32899 Scan Post
    Instrument
    • SM.05 PDF p. 188 (printed 179) · Digital solutions, intraoral scanT0 25251 Hex Driver (Manua[l])
    Torque

    D013 section 15.1 step 3 (PDF p. 4) names the M1.6 screw and a HEX wrench but gives no torque. Confirm with HealthIN.

  2. 2Model

    Digital implant analogue

    For a printed model. The M2 screw fixes the analogue to the model.

    Cited Catalogue 2026 PDF p. 28 (printed 27); SM.05 PDF p. 91 (printed 82)

    Screw

    Model fixation only; not an implant or occlusal screw.

    • Catalogue 2026 PDF p. 28 (printed 27)
    • SM.05 PDF p. 91 (printed 82)
    Instrument

    No instrument reference is given for this step. Confirm with HealthIN.

    Torque

    Not stated in the cited IFU. Confirm with HealthIN.

  3. 3Final

    CAD/CAM Ti-Base, NO-HEX

    NO-HEX references are documented for multiple-tooth restorations. Choose GH 0.7, 1.5 or 2.4 mm.

    Cited Catalogue 2026 PDF p. 27 (printed 26); SM.05 PDF p. 183, 189 (printed 174, 180)

    Screw
    • SM.05 PDF p. 184 (printed 175) · CAD/CAM Ti-base abutment tableAbutment Screw T6 16000
    Instrument

    Hand-tighten with the manual HEX driver, then torque with the ratchet and a short or long ratchet HEX driver.

    • SM.05 PDF p. 203 (printed 194) · “Tightening – Torque Values” table, row “Cad/Cam Ti-Base Abutment Screws”
    Torque

    Max. 30 Ncm

    • D013 PDF p. 2 · section 7 torque table, row “CAD-CAM Ti-Base Abutment Screws”Max. 30 Ncm
    • D013 PDF p. 5 · section 15.1, step 13torqued up to a maximum of 30 Ncm
    • SM.05 PDF p. 190 (printed 181) · Bridge, step 6torqued to a maximum of 30 Ncm with a ratchet
  4. 4Final

    Angled screw-channel Ti-Base, NO-HEX

    Where the design uses an angled screw channel. GH 1, 1.5 or 2.5 mm.

    Cited Catalogue 2026 PDF p. 27 (printed 26); D06.001/EN PDF p. 8, 9, 14, 15 (printed 8-9, 14-15)

    Screw

    Match the GH: 1 mm → T6 16011, 1.5 mm → T6 16014, 2.5 mm → T6 16012. Not interchangeable.

    • SM.05 PDF p. 184 (printed 175) · CAD/CAM Ti-base abutment tableAbutment Screws T6 16011-T6 16014-T6 16012
    Instrument
    • T0 2027 Angled ratchet tightening instrument

    The manual calls for an angled HEX driver without a reference. The Prosthetic Driver Kit contains T0 2027. Confirm with HealthIN.

    • SM.05 PDF p. 193 (printed 184) · Total edentulism with CAD/CAM Ti-Base, step 5tightened with the angled hex driver
    Torque

    Sources differ

    • Max. 30 Ncm D013 PDF p. 2 · section 7 torque table, row “CAD-CAM Ti-Base Abutment Screws”Max. 30 Ncm
    • Max. 25 Ncm SM.05 PDF p. 193 (printed 184) · Total edentulism with CAD/CAM Ti-Base, step 6angled ti-base abutments must be torqued to a maximum of 25 Ncm

    Confirm with HealthIN before tightening.

Confirm with HealthIN

  • Torque and driver for angled screw-channel Ti-Base screws (sources differ).
  • HEX Ti-Bases (T6 32804/32823/32806, T6 32817/32825/32818) are documented for single teeth only; not for a connected bridge.
  • Scan body screw torque and digital-analogue instrument (not stated in D013).

Documents and editions

  • D013 CAD/CAM abutments — instructions for use · D013 Rev 000 · published 24 Oct 2025 · Open D013 PDF · https://nucleoss.co.uk/documents/cadcam-abutments-ifu.pdf
  • Catalogue 2026 NucleOSS 2026 product catalogue · Rev 05 · March 2026 · Open Catalogue 2026 PDF · https://nucleoss.co.uk/documents/nucleoss-2026-catalogue.pdf
  • D06.001/EN Digital library selection guide · D06.001/EN Rev 01 · 18 Aug 2022 · Open D06.001/EN PDF · https://nucleoss.co.uk/documents/digital-library-selection-guide.pdf
  • SM.05 NucleOSS prosthetic manual · Rev00 · 15 June 2026 · manufacturer document held by HealthIN, not published on this site

Sources checked 11 October 2026. Check the edition against the document supplied with your components.