NucleOSS workflow sheet · Full-arch or multi-unit bridge, digital workflow

Full arch on multi-unit abutments

NO-HEX multi-unit abutments (0°, 17°, 30°), scan post or ScanWings, multi-unit analogue and Ti-Base with occlusal screw.

A parts reference for dental laboratories and dentists, compiled from the manufacturer documents listed below. It is not a clinical recommendation: the clinician selects the components, and the current instructions for use apply. Torque is shown only where an IFU states it. Confirm with HealthIN marks anything the cited sources do not state or state differently.

To print or save this sheet as a PDF, use your browser’s Print command.

Ask HealthIN about this sheet
  1. 1Abutment

    Multi-unit abutment 0°, NO-HEX

    GH 1, 2.5 or 4 mm.

    Cited Catalogue 2026 PDF p. 21 (printed 20); SM.05 PDF p. 128 (printed 119)

    Screw

    None listed

    No separate screw is listed for the straight base; T6 16008 is documented for angled bases only.

    Instrument
    • SM.05 PDF p. 128 (printed 119) · Multi-unit abutment placementScrew-Retained Abutment Tightening Part T0 25206
    • SM.05 PDF p. 204 (printed 195) · “Tightening – Torque Values” table, row “Multi Unit Straight Abutments”
    Torque

    Max. 30 Ncm

    • D010 PDF p. 2 · section 7 torque table, row “Multi Unit Bases”Max. 30 Ncm
  2. 2Abutment

    Multi-unit abutment 17° or 30°, NO-HEX

    17°: T6 32425–32427 · 30°: T6 32428–32430 (GH 1.5, 2.5, 4 mm).

    Cited Catalogue 2026 PDF p. 21 (printed 20); SM.05 PDF p. 128 (printed 119)

    Screw
    • SM.05 PDF p. 128 (printed 119) · Multi-unit angled abutmentfixed to the implant with the abutment screw included in the package
    Instrument
    • SM.05 PDF p. 204 (printed 195) · “Tightening – Torque Values” table, row “Multi Unit Angled (17° / 30°) Abutment Screws”
    Torque

    Max. 30 Ncm

    • D010 PDF p. 2 · section 7 torque table, row “Multi Unit Bases”Max. 30 Ncm
  3. 3Scan

    Multi-unit scan post, NO-HEX

    Cited Catalogue 2026 PDF p. 22, 28 (printed 21, 27); SM.05 PDF p. 181, 197 (printed 172, 188)

    Screw
    • SM.05 PDF p. 198 (printed 189) · Multi-unit CAD/CAM Ti-Base, intraoral scanT6 32039 Scan Post · T6 16009 Occlusal Screw
    Instrument
    • SM.05 PDF p. 198 (printed 189) · Multi-unit CAD/CAM Ti-Base, intraoral scanT0 25251 Hex Driver (Manual)
    Torque

    Sources differ

    • Max. 5 Ncm D013 PDF p. 5 · section 15.2, intraoral scan, step 2The screw of the scanning part is torqued with a maximum of 5 Ncm
    • Max. 10 Ncm SM.05 PDF p. 198 (printed 189) · Multi-unit CAD/CAM Ti-Base, intraoral scanmust be torqued to a maximum of 10 Ncm

    Confirm with HealthIN. The manual warns that over-tightening can distort the scan post.

  4. 4Scan

    ScanWings full-arch scan posts (alternative)

    Kit YRS-0033: scan post T6 32040 with extensions T0 4000-1, -2, -3.

    Cited Catalogue 2026 PDF p. 50 (printed 49); D06.001/EN PDF p. 7, 13

    Screw
    • Catalogue 2026 PDF p. 50 (printed 49)
    Instrument
    • Catalogue 2026 PDF p. 50 (printed 49)
    Torque

    Not stated in the cited IFU. Confirm with HealthIN.

  5. 5Model

    Multi-unit digital analogue

    Cited Catalogue 2026 PDF p. 28 (printed 27); SM.05 PDF p. 91, 147, 199 (printed 82, 138, 190); SM.05 PDF p. 91 (printed 82)

    Screw

    Model fixation only.

    • SM.05 PDF p. 199 (printed 190) · Preparation of 3D dental modelT0 2015 Digital Analog Screw
    Instrument
    • D013 PDF p. 5 · section 15.2, 3D model, step 3with the help of a manual hex driver
    Torque

    Not stated in the cited IFU. Confirm with HealthIN.

  6. 6Final

    Multi-unit Ti-Base, NO-HEX

    Physical length L 6.5 or 4.5 mm and the library subfile are not given in the catalogue. Confirm with HealthIN.

    Cited Catalogue 2026 PDF p. 22, 28 (printed 21, 27); SM.05 PDF p. 197, 199 (printed 188, 190)

    Screw
    • SM.05 PDF p. 197 (printed 188) · Multi-unit CAD/CAM componentsT6 16009 Occlusal Screw
    Instrument
    • SM.05 PDF p. 204 (printed 195) · “Tightening – Torque Values” table, row “Occlusal Screws Attached to Prosthesis”
    Torque

    Max. 20 Ncm

    • D010 PDF p. 2 · section 7 torque table, row “Abutment Occlusal Screw”20 Ncm
    • D010 PDF p. 6 · section 15.4, step 2torqued to a maximum of 20 Ncm
    • D013 PDF p. 6 · section 15.2, placing the bridge, step 2maximum torque value of 20 Ncm

Confirm with HealthIN

  • Scan post torque (D013 and the manual differ) and ScanWings screw torque.
  • Multi-unit Ti-Base length (L 6.5 or 4.5 mm) and library subfile; library mapping for T6 32430.
  • HEX multi-unit parts (T6 32413–32415, T6 32038, T6 32814) are documented for single teeth only.

Documents and editions

  • D010 Multi-unit abutments — instructions for use · D010 Rev 000 · published 24 Oct 2025 · Open D010 PDF · https://nucleoss.co.uk/documents/multi-unit-abutments-ifu.pdf
  • D013 CAD/CAM abutments — instructions for use · D013 Rev 000 · published 24 Oct 2025 · Open D013 PDF · https://nucleoss.co.uk/documents/cadcam-abutments-ifu.pdf
  • Catalogue 2026 NucleOSS 2026 product catalogue · Rev 05 · March 2026 · Open Catalogue 2026 PDF · https://nucleoss.co.uk/documents/nucleoss-2026-catalogue.pdf
  • D06.001/EN Digital library selection guide · D06.001/EN Rev 01 · 18 Aug 2022 · Open D06.001/EN PDF · https://nucleoss.co.uk/documents/digital-library-selection-guide.pdf
  • SM.05 NucleOSS prosthetic manual · Rev00 · 15 June 2026 · manufacturer document held by HealthIN, not published on this site

Sources checked 11 October 2026. Check the edition against the document supplied with your components.