NucleOSS workflow sheet · Healing phase before the restoration
Healing: cover screws and gingiva formers
Cover screws by implant diameter, cylindrical, platform-switch, PEEK and Ti-Base gingiva formers.
A parts reference for dental laboratories and dentists, compiled from the manufacturer documents listed below. It is not a clinical recommendation: the clinician selects the components, and the current instructions for use apply. Torque is shown only where an IFU states it. Confirm with HealthIN marks anything the cited sources do not state or state differently.
To print or save this sheet as a PDF, use your browser’s Print command.
Ask HealthIN about this sheet-
1Cover
Implant cover screw
Match the implant: Ø 3.5 T6 32150 · Ø 4.0 T6 32158 · Ø 4.5 T6 32157 · Ø 5.0 T6 32151 · Ø 6.0 T6 32152.
- T6 32150 Implant cover screw · for T6 Torq Ø 3.5 mm implants
- T6 32158 Implant cover screw · for T6 Torq Ø 4.0 mm implants
- T6 32157 Implant cover screw · for T6 Torq Ø 4.5 mm implants
- T6 32151 Implant cover screw · for T6 Torq Ø 5.0 mm implants
- T6 32152 Implant cover screw · for T6 Torq Ø 6.0 mm implants
Cited Catalogue 2026 PDF p. 12, 14 (printed 11, 13)
- Screw
-
None listed
No separate screw is listed. The catalogue shows the cover screw in the implant packaging.
- Instrument
-
- T0 25251 Manual HEX driver
- T0 2024 Short ratchet HEX driver
- T0 2025 Long ratchet HEX driver
- T0 25900 Torque ratchet
- SM.05 PDF p. 203 (printed 194) · “Tightening – Torque Values” table, row “Implant Cover Screw And Sinus Caps”
- Torque
-
By hand or ratchet (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
By Hand/Raschette (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
-
2Healing
Cylindrical gingiva former
Ø 4.3: T6 NR512–NR514 · Ø 5.0: T6 SD515–SD517 · Ø 6.0: T6 WD518–WD520 (GH 2, 4, 6 mm).
- T6 NR512 Ø 4.3 mm platform · GH 2 mm
- T6 NR513 Ø 4.3 mm platform · GH 4 mm
- T6 NR514 Ø 4.3 mm platform · GH 6 mm
- T6 SD515 Ø 5.0 mm platform · GH 2 mm
- T6 SD516 Ø 5.0 mm platform · GH 4 mm
- T6 SD517 Ø 5.0 mm platform · GH 6 mm
- T6 WD518 Ø 6.0 mm platform · GH 2 mm
- T6 WD519 Ø 6.0 mm platform · GH 4 mm
- T6 WD520 Ø 6.0 mm platform · GH 6 mm
Cited Catalogue 2026 PDF p. 16 (printed 15); SM.05 PDF p. 39, 42, 44 (printed 30, 33, 35)
- Screw
-
None listed
No separate screw is listed.
- Instrument
-
- T0 25251 Manual HEX driver
- T0 2024 Short ratchet HEX driver
- T0 2025 Long ratchet HEX driver
- T0 25900 Torque ratchet
- SM.05 PDF p. 203 (printed 194) · “Tightening – Torque Values” table, row “Gingiva Former”
- Torque
-
By hand or ratchet (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
By Hand/Raschette (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
-
3Healing
Platform-switch gingiva former
Ø 4.3: T6 NR501–NR503 · Ø 5.0: T6 SD504–SD506 · Ø 6.0: T6 WD507–WD509 (GH 2, 4, 6 mm).
- T6 NR501 Ø 4.3 mm platform · GH 2 mm
- T6 NR502 Ø 4.3 mm platform · GH 4 mm
- T6 NR503 Ø 4.3 mm platform · GH 6 mm
- T6 SD504 Ø 5.0 mm platform · GH 2 mm
- T6 SD505 Ø 5.0 mm platform · GH 4 mm
- T6 SD506 Ø 5.0 mm platform · GH 6 mm
- T6 WD507 Ø 6.0 mm platform · GH 2 mm
- T6 WD508 Ø 6.0 mm platform · GH 4 mm
- T6 WD509 Ø 6.0 mm platform · GH 6 mm
Cited Catalogue 2026 PDF p. 16 (printed 15); SM.05 PDF p. 39, 42, 44 (printed 30, 33, 35)
- Screw
-
None listed
No separate screw is listed.
- Instrument
-
- T0 25251 Manual HEX driver
- T0 2024 Short ratchet HEX driver
- T0 2025 Long ratchet HEX driver
- T0 25900 Torque ratchet
- SM.05 PDF p. 203 (printed 194) · “Tightening – Torque Values” table, row “Gingiva Former”
- Torque
-
By hand or ratchet (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
By Hand/Raschette (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
-
4Healing
PEEK gingiva former
Shapeable PEEK former; supplied with its abutment screw.
- T6 32034 PEEK gingiva former
Cited Catalogue 2026 PDF p. 16 (printed 15); SM.05 PDF p. 44 (printed 35)
- Screw
-
- T6 16000 Abutment screw
- SM.05 PDF p. 44 (printed 35) · Temporary gingiva former, package content
supplied in the same package with abutment screw
- Instrument
-
- T0 25251 Manual HEX driver
- T0 2024 Short ratchet HEX driver
- T0 2025 Long ratchet HEX driver
- T0 25900 Torque ratchet
- SM.05 PDF p. 203 (printed 194) · “Tightening – Torque Values” table, row “Connection Screws of Peek Parts”
- Torque
-
10 Ncm
- D007 PDF p. 2 · section 7 torque table, row “Temporary (PEEK) Parts and Screws”
10 Ncm
- D007 PDF p. 2 · section 7 torque table, row “Temporary (PEEK) Parts and Screws”
-
5Healing
Ti-Base gingiva former
For digital Ti-Base workflows. Ø 4.5 mm, GH 4 or 6 mm.
Cited Catalogue 2026 PDF p. 28 (printed 27); SM.05 PDF p. 188 (printed 179)
- Screw
-
None listed
No separate screw is listed.
- Instrument
-
- SM.05 PDF p. 179 (printed 170) · Placement of Ti-base gingiva former
T0 25251 Hex Driver (Manua[l]) · T0 2024 Tightening Driver (short)
- SM.05 PDF p. 179 (printed 170) · Placement of Ti-base gingiva former
- Torque
-
By hand or ratchet (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
By Hand/Raschette (10 Ncm)
- D007 PDF p. 2 · section 7 torque table, row “All Covers and Healing Heads”
Confirm with HealthIN
- Whether a separate cover screw is needed (the catalogue shows one in the implant packaging).
- Gingiva-former GH and platform for each site.
Documents and editions
- D007 Straight abutments — instructions for use · D007 Rev 000 · published 24 Oct 2025 · Open D007 PDF · https://nucleoss.co.uk/documents/straight-abutments-ifu.pdf
- Catalogue 2026 NucleOSS 2026 product catalogue · Rev 05 · March 2026 · Open Catalogue 2026 PDF · https://nucleoss.co.uk/documents/nucleoss-2026-catalogue.pdf
- SM.05 NucleOSS prosthetic manual · Rev00 · 15 June 2026 · manufacturer document held by HealthIN, not published on this site
Sources checked 11 October 2026. Check the edition against the document supplied with your components.