NucleOSS workflow sheet · Single crown, digital workflow, implant level

Single screw-retained crown on a CAD/CAM Ti-Base

Scan body, digital analogue and a HEX CAD/CAM Ti-Base with its abutment screw, driver and IFU torque.

A parts reference for dental laboratories and dentists, compiled from the manufacturer documents listed below. It is not a clinical recommendation: the clinician selects the components, and the current instructions for use apply. Torque is shown only where an IFU states it. Confirm with HealthIN marks anything the cited sources do not state or state differently.

To print or save this sheet as a PDF, use your browser’s Print command.

Ask HealthIN about this sheet
  1. 1Scan

    Implant-level scan body

    One per implant. T6 32895 is the short version.

    Cited Catalogue 2026 PDF p. 27 (printed 26); D06.001/EN PDF p. 2, 3 (printed 2-3)

    Screw

    The manual names T6 16000 with T6 32899; confirm the screw for the short T6 32895 with HealthIN.

    • SM.05 PDF p. 188 (printed 179) · Digital solutions, intraoral scanT6 16000 Abutment Screw · T6 32899 Scan Post
    Instrument
    • SM.05 PDF p. 188 (printed 179) · Digital solutions, intraoral scanT0 25251 Hex Driver (Manua[l])
    Torque

    D013 section 15.1 step 3 (PDF p. 4) names the M1.6 screw and a HEX wrench but gives no torque. Confirm with HealthIN.

  2. 2Model

    Digital implant analogue

    For a printed model. The M2 screw fixes the analogue to the model.

    Cited Catalogue 2026 PDF p. 28 (printed 27); SM.05 PDF p. 91 (printed 82)

    Screw

    Model fixation only; not an implant or occlusal screw.

    • Catalogue 2026 PDF p. 28 (printed 27)
    • SM.05 PDF p. 91 (printed 82)
    Instrument

    No instrument reference is given for this step. Confirm with HealthIN.

    Torque

    Not stated in the cited IFU. Confirm with HealthIN.

  3. 3Final

    CAD/CAM Ti-Base, HEX

    HEX references are documented for single-tooth restorations. Choose GH 0.7, 1.5 or 2.4 mm.

    Cited Catalogue 2026 PDF p. 27 (printed 26); SM.05 PDF p. 183 (printed 174)

    Screw
    • SM.05 PDF p. 184 (printed 175) · CAD/CAM Ti-base abutment tableAbutment Screw T6 16000
    Instrument

    Hand-tighten with the manual HEX driver, then torque with the ratchet and a short or long ratchet HEX driver.

    • SM.05 PDF p. 203 (printed 194) · “Tightening – Torque Values” table, row “Cad/Cam Ti-Base Abutment Screws”
    Torque

    Max. 30 Ncm

    • D013 PDF p. 2 · section 7 torque table, row “CAD-CAM Ti-Base Abutment Screws”Max. 30 Ncm
    • D013 PDF p. 5 · section 15.1, step 13torqued up to a maximum of 30 Ncm

Confirm with HealthIN

  • The screw for the short scan body T6 32895.
  • Scan body screw torque and digital-analogue instrument (not stated in D013).
  • Current digital-library file and version for your design software.

Documents and editions

  • D013 CAD/CAM abutments — instructions for use · D013 Rev 000 · published 24 Oct 2025 · Open D013 PDF · https://nucleoss.co.uk/documents/cadcam-abutments-ifu.pdf
  • Catalogue 2026 NucleOSS 2026 product catalogue · Rev 05 · March 2026 · Open Catalogue 2026 PDF · https://nucleoss.co.uk/documents/nucleoss-2026-catalogue.pdf
  • D06.001/EN Digital library selection guide · D06.001/EN Rev 01 · 18 Aug 2022 · Open D06.001/EN PDF · https://nucleoss.co.uk/documents/digital-library-selection-guide.pdf
  • SM.05 NucleOSS prosthetic manual · Rev00 · 15 June 2026 · manufacturer document held by HealthIN, not published on this site

Sources checked 11 October 2026. Check the edition against the document supplied with your components.